Welcome to the IKCEST

Journal of Synchrotron Radiation | Vol., Issue. | 2020-04-14 | Pages

Journal of Synchrotron Radiation

Measurement of the horizontal beam emittance of undulator radiation by tandem‐double‐slit optical system

Takumi Ikeda   Motoki Kawashima   Yasushi Kagoshima   Yuki Aoi   Tatsuki Akada   Yuki Takayama  
Abstract

A tandem‐double‐slit optical system was constructed to evaluate the practical beam emittance of undulator radiation. The optical system was a combination of an upstream slit (S1) and downstream slit (S2) aligned on the optical axis with an appropriate separation. The intensity distribution after the double slits, I(x1, x2), was measured by scanning S1 and S2 in the horizontal direction. Coordinates having intensity were extracted from I(x1, x2), whose contour provided the standard deviation ellipse in the x1–x2 space. I(x1, x2) was converted to the corresponding distribution in the phase space, I(x1, x1′). The horizontal beam emittance was evaluated to be 3.1 nm rad, which was larger than the value of 2.4 nm rad estimated by using ray‐tracing. It was found that the increase was mainly due to an increase in beam divergence rather than size.

Original Text (This is the original text for your reference.)

Measurement of the horizontal beam emittance of undulator radiation by tandem‐double‐slit optical system

A tandem‐double‐slit optical system was constructed to evaluate the practical beam emittance of undulator radiation. The optical system was a combination of an upstream slit (S1) and downstream slit (S2) aligned on the optical axis with an appropriate separation. The intensity distribution after the double slits, I(x1, x2), was measured by scanning S1 and S2 in the horizontal direction. Coordinates having intensity were extracted from I(x1, x2), whose contour provided the standard deviation ellipse in the x1–x2 space. I(x1, x2) was converted to the corresponding distribution in the phase space, I(x1, x1′). The horizontal beam emittance was evaluated to be 3.1 nm rad, which was larger than the value of 2.4 nm rad estimated by using ray‐tracing. It was found that the increase was mainly due to an increase in beam divergence rather than size.

+More

Cite this article
APA

APA

MLA

Chicago

Takumi Ikeda, Motoki Kawashima,Yasushi Kagoshima, Yuki Aoi, Tatsuki Akada, Yuki Takayama,.Measurement of the horizontal beam emittance of undulator radiation by tandem‐double‐slit optical system. (),.

Disclaimer: The translated content is provided by third-party translation service providers, and IKCEST shall not assume any responsibility for the accuracy and legality of the content.
Translate engine
Article's language
English
中文
Pусск
Français
Español
العربية
Português
Kikongo
Dutch
kiswahili
هَوُسَ
IsiZulu
Action
Recommended articles

Report

Select your report category*



Reason*



By pressing send, your feedback will be used to improve IKCEST. Your privacy will be protected.

Submit