Rapid gas chromatographic determination of ethylene oxide, ethylene chlorohydrin, and ethylene glycol residues in rubber catheters.

R J, Muzeni  
Abstract

Isothermal gas chromatography with flame ionization detection was used to determine residual ethylene oxide (EtO), ethylene chlorohydrin, and ethylene glycol in soft rubber catheters that had been sterilized with EtO. Catheter samples were extracted by shaking with carbon disulfide, and the extract was analyzed on a 3% Carbowax 20M on 80-100 mesh Chromosorb 101 column, using nitrogen as the carrier gas. Ten replicate injections of a mixed standards solution gave coefficients of variation of 1.91, 1.23, and 4.74% for EtO, ethylene chlorohydrin, and ethylene glycol, respectively. A linear response was obtained with concentrations ranging from 1.0 to 7.9 micrograms EtO, 14.0 to 88.0 micrograms ethylene chlorohydrin, and 31.0 to 98.5 micrograms ethylene glycol. The proposed method detected as little as 0.5, 5.0, and 16.5 ng EtO, ethylene chlorohydrin, and ethylene glycol, respectively.

Original Text (This is the original text for your reference.)

Rapid gas chromatographic determination of ethylene oxide, ethylene chlorohydrin, and ethylene glycol residues in rubber catheters.

Isothermal gas chromatography with flame ionization detection was used to determine residual ethylene oxide (EtO), ethylene chlorohydrin, and ethylene glycol in soft rubber catheters that had been sterilized with EtO. Catheter samples were extracted by shaking with carbon disulfide, and the extract was analyzed on a 3% Carbowax 20M on 80-100 mesh Chromosorb 101 column, using nitrogen as the carrier gas. Ten replicate injections of a mixed standards solution gave coefficients of variation of 1.91, 1.23, and 4.74% for EtO, ethylene chlorohydrin, and ethylene glycol, respectively. A linear response was obtained with concentrations ranging from 1.0 to 7.9 micrograms EtO, 14.0 to 88.0 micrograms ethylene chlorohydrin, and 31.0 to 98.5 micrograms ethylene glycol. The proposed method detected as little as 0.5, 5.0, and 16.5 ng EtO, ethylene chlorohydrin, and ethylene glycol, respectively.

+More

Cite this article
APA

APA

MLA

Chicago

R J, Muzeni,.Rapid gas chromatographic determination of ethylene oxide, ethylene chlorohydrin, and ethylene glycol residues in rubber catheters.. 68 (3),506-8.

Disclaimer: The translated content is provided by third-party translation service providers, and IKCEST shall not assume any responsibility for the accuracy and legality of the content.
Translate engine
Article's language
English
中文
Pусск
Français
Español
العربية
Português
Kikongo
Dutch
kiswahili
هَوُسَ
IsiZulu
Action
Recommended articles

Report

Select your report category*



Reason*



By pressing send, your feedback will be used to improve IKCEST. Your privacy will be protected.

Submit
Cancel